Key result
Electronic stethoscope with spectral analysis detects severe carotid stenosis with an AUC of ~0.79.
Why the study?
Stroke prevention needs accessible carotid stenosis screening tools, but acoustic stethoscopes are limited by subjectivity and inconsistent diagnostic accuracy.
Does spectral analysis of electronic stethoscope recordings accurately detect severe carotid stenosis and plaques compared to duplex ultrasound in patients with suspected carotid disease?
Cross-Sectional (n=94)
No
Does spectral analysis of electronic stethoscope recordings accurately detect severe carotid stenosis and plaques compared to duplex ultrasound in patients with suspected carotid disease?
An electronic stethoscope equipped with spectral analysis offers a promising and accessible initial screening tool for identifying severe carotid stenosis, outperforming traditional auscultation.
Moderate accuracy for severe carotid stenosis; hypothesis-generating and requires prospective validation before screening use.
Background: Stroke is the second leading cause of death worldwide, with carotid stenosis being a primary contributor. Therefore, stroke prevention would benefit from accessible carotid stenosis screening tools. Historically, acoustic stethoscopes were used to listen to the carotid artery, but this method is now outdated due to its subjectivity and inconsistent sensitivity and specificity in detecting stenosis. In contrast, electronic stethoscopes record audio, enabling precise and objective analysis. To overcome traditional auscultation limitations, our study introduces a signal analysis scheme to evaluate the electronic stethoscope as a potential screening tool for carotid plaques and severe stenosis. Methods: We included 94 patients undergoing duplex ultrasound (DUS) for recent transient ischemic attack (TIA) or pre-operative assessment for carotid endarterectomy. DUS served as the clinical reference for determining plaque presence and estimating carotid stenosis. Participants held their breath during electronic stethoscope measurements at two points along each carotid artery: (I) proximal, on the common carotid; and (II) distal, near the bifurcation. From these recordings, we extracted 10 spectral features and utilized multivariable binary logistic regression for predicting plaques and severe stenosis, applying 10-fold cross-validation for internal validation. We constructed the receiver operating characteristic (ROC) curve by plotting the true positive rate against the false positive rate at various cutoff settings. We reported the area under the curve (AUC), along with sensitivity and specificity, which were determined using a single optimal cutoff point. Results: For detecting >70% stenosis using distal location recordings, the analysis yielded training and testing AUCs of 0.87 and 0.79, sensitivity of 84.9% and 78.6%, and specificity of 73.6% and 72.1%, respectively. Using proximal location recordings, training and testing AUCs were 0.84 and 0.73, with sensitivities of 79.8% and 60.7%, and specificities of 76.0% and 75.6%, respectively. For detecting the presence of plaques, proximal location measurements showed training and testing AUCs of 0.79 and 0.7, sensitivities of 54.9% and 51.9%, and specificities of 91.9% and 78.8%, respectively. Conclusions: Our findings demonstrate that the electronic stethoscope with spectral analysis is promising for identifying severe stenosis but has limited sensitivity for detecting any plaque. The performance obtained with this approach is superior to that attainable with conventional auscultation. This approach could serve as a promising, user-friendly screening tool, particularly in resource-limited settings.
No takes yet. Share an insight, caveat, or question.
Malik et al. (2025) conducted a cross-sectional in Carotid stenosis (n=94). Electronic stethoscope with spectral analysis vs. Duplex ultrasound (DUS) was evaluated on Detection of severe (>70%) carotid stenosis using distal location recordings (AUC 0.79). An electronic stethoscope equipped with spectral bruit analysis detected severe carotid stenosis with a testing area under the curve of 0.79, a sensitivity of 78.6%, and a specificity of 72.1%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: