September 16, 2009IEEE Transactions on Biomedical Engineering151 citationsOpen Access

Robust Detection of Premature Ventricular Contractions Using a Wave-Based Bayesian Framework

OSOmid SayadiMSMohammad Bagher ShamsollahiGCGari D. Clifford

Key Result

A wave-based Bayesian framework algorithm achieved an average detection accuracy of 99.10%, aggregate sensitivity of 98.77%, and aggregate positive predictivity of 97.47% for detecting PVCs.

Structured PICO

P
Population
Algorithm validation study using 5 databases, including 40 records from the MIT-BIH arrhythmia database, to detect premature ventricular contractions.
I
Intervention
Model-based dynamic algorithm for tracking ECG characteristic waveforms using an extended Kalman filter and a polargram
O
Outcome
Detection accuracy, sensitivity, and positive predictivity for differentiating normal, PVC, and other beatssurrogate

A novel wave-based Bayesian framework using an extended Kalman filter and polargram achieves high accuracy (>99%) in detecting premature ventricular contractions from ECG signals.

Abstract

Detection and classification of ventricular complexes from the ECG is of considerable importance in Holter and critical care patient monitoring, being essential for the timely diagnosis of dangerous heart conditions. Accurate detection of premature ventricular contractions (PVCs) is particularly important in relation to life-threatening arrhythmias. In this paper, we introduce a model-based dynamic algorithm for tracking the ECG characteristic waveforms using an extended Kalman filter. The algorithm can work on single or multiple leads. A "polargram"--a polar representation of the signal--is introduced, which is constructed using the Bayesian estimations of the state variables. The polargram allows the specification of a polar envelope for normal rhythms. Moreover, we propose a novel measure of signal fidelity by monitoring the covariance matrix of the innovation signals throughout the filtering procedure. PVCs are detected by simultaneous tracking the signal fidelity and the polar envelope. Five databases, including 40 records from MIT-BIH arrhythmia database, are used for differentiating normal, PVC, and other beats. Performance evaluation results show that the proposed method has an average detection accuracy of 99.10%, aggregate sensitivity of 98.77%, and aggregate positive predictivity of 97.47%. Furthermore, the method is capable of 100% accuracy for records that contain only PVCs and normal sinus beats. The results illustrate that the method can contribute to, and enhance the performance of clinical PVC detection.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Sayadi et al. (2009) studied Premature ventricular contractions. Wave-based Bayesian framework algorithm was evaluated on Average detection accuracy. A wave-based Bayesian framework algorithm achieved an average detection accuracy of 99.10%, aggregate sensitivity of 98.77%, and aggregate positive predictivity of 97.47% for detecting PVCs.

synapsesocial.com/papers/6a3144c29750ae3d276ea998https://doi.org/10.1109/tbme.2009.2031243
Ask AI
Helpful
Bookmark
Share
View Full Paper