Key result
Contactless laser Doppler vibrometry matches pulsed-wave Doppler echo for LVET measurements within ~5 ms.
Why the study?
Routine use of LVET in first-line population screening is hindered by the operational and time demands of operator-dependent modalities like echocardiography.
Does contactless suprasternal notch laser Doppler vibrometry accurately and automatically measure left ventricular ejection time compared to pulsed-wave Doppler echocardiography?
Cross-Sectional (n=238)
Does contactless suprasternal notch laser Doppler vibrometry accurately and automatically measure left ventricular ejection time compared to pulsed-wave Doppler echocardiography?
Mean Difference: -4.75
Absolute Event Rate: 304% vs 309%
p-value: p=0.0 for equivalence
Contactless suprasternal notch laser vibrometry combined with deep learning provides a rapid, accurate, and operator-independent method for measuring left ventricular ejection time, potentially enabling large-scale cardiovascular screening.
May enable contactless LVET assessment; leaves open need for prospective validation before clinical use.
Left Ventricular Ejection Time (LVET) is a sensitive indicator of early cardiac dysfunction, yet its routine use in first-line population screening is hindered by the operational and time demands of operator-dependent modalities like echocardiography. In this study, we introduce a scalable, contactless approach using laser Doppler vibrometry (LDV) signals targeted at the suprasternal notch (SSN) to capture hemodynamic events without physical sensor attachment. We demonstrate in a cohort of 238 individuals that this rapid, zero-setup measurement yields LVET values statistically equivalent to pulsed-wave Doppler echocardiography (concordance within a 25 ms margin). Furthermore, an exploratory clinical analysis revealed that the LDV-derived LVET index (LVETI) was significantly prolonged in patients with aortic insufficiency and stenosis compared to low-risk controls, demonstrating the clinical plausibility of the measurement. To eliminate the bottleneck of expert interpretation, we developed and validated a deep learning model that automatically extracts LVET with high accuracy (mean absolute error of 8.08 ms). By combining easy LDV data acquisition with automated analysis, this approach may facilitate rapid, operator-independent cardiovascular assessment, potentially enhancing population-level screening and cardiovascular health management.
No takes yet. Share an insight, caveat, or question.
Gonzalez-Billandon et al. (2026) conducted a cross-sectional in Cardiovascular disease screening (n=238). Contactless suprasternal notch laser Doppler vibrometry vs. Pulsed-wave Doppler echocardiography was evaluated on Left Ventricular Ejection Time (LVET) (MD -4.75 ms, p=0.0 for equivalence). Contactless suprasternal notch laser Doppler vibrometry yielded left ventricular ejection time measurements statistically equivalent to pulsed-wave Doppler echocardiography within a 25 ms margin (mean difference -4.75 ms).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: