Key result
New contact model successfully simulates bio-prosthetic valve coaptation and reveals asymmetrical fibers cause swirling motion.
Why the study?
A new contact model was needed to prevent inter-leaflet penetration of bio-prosthetic heart valves at end-systole by directly applying kinematic constraints and creating smooth free edges.
A novel contact model based on the coefficient of restitution improves the accuracy of bio-prosthetic heart valve simulations, capturing realistic leaflet dynamics and coaptation.
Improves bioprosthetic valve simulations; leaves open clinical durability benefits pending validation.
A new general contact model is proposed for preventing inter-leaflet penetration of bio-prosthetic heart valves (BHV) at the end of the systole, which has the advantage of applying kinematic constraints directly and creating smooth free edges. At the end of each time step, the impenetrability constraints and momentum exchange between the impacting bodies are applied separately based on the coefficient of restitution. The contact method is implemented in a rotation-free, large deformation, and thin shell finite-element (FE) framework based on loop's subdivision surfaces. A nonlinear, anisotropic material model for a BHV is employed which uses Fung-elastic constitutive laws for in-plane and bending responses, respectively. The contact model is verified and validated against several benchmark problems. For a BHV-specific validation, the computed strains on different regions of a BHV under constant pressure are compared with experimentally measured data. Finally, dynamic simulations of BHV under physiological pressure waveform are performed for symmetrical and asymmetrical fiber orientations incorporating the new contact model and compared with the penalty contact method. The proposed contact model provides the coaptation area of a functioning BHV during the closing phase for both of the fiber orientations. Our results show that fiber orientation affects the dynamic of leaflets during the opening and closing phases. A swirling motion for the BHV with asymmetrical fiber orientation is observed, similar to experimental data. To include the fluid effects, fluid-structure interaction (FSI) simulation of the BHV is performed and compared to the dynamic results.
No takes yet. Share an insight, caveat, or question.
Asadi et al. (2023) studied Bio-prosthetic heart valve simulation. Contact model based on the coefficient of restitution vs. Penalty contact method was evaluated on Coaptation area and dynamic of leaflets during opening and closing phases. A new contact model based on the coefficient of restitution successfully simulated bio-prosthetic heart valve coaptation and showed that asymmetrical fiber orientation causes a swirling motion.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: