Demonstrates glutathione detection in serum using a novel signal-on sensor, implying diagnostic potential.
NPs/PPy-PL/GCE gradually recovered upon incubation with the linearly increased concentration of GSH from 5 nM to 5 μM, and the calculated LOD and LOQ were 0.12 nM and 0.36 nM, respectively. The proposed sensor displays an excellent anti-interference ability and successfully demonstrates the sensing of GSH in human serum samples, which shows a good recovery percentage from 102% to 104%. The superior sensing performance of the proposed signal-on solid-state ECL-RET sensor methodology provides a promising potential for clinical diagnostics.
No takes yet. Share an insight, caveat, or question.
Paramasivam et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: