This paper reports an approach for the direct determination of glutathione in individual Ramos cells based upon capillary electrophoresis coupled with electrogenerated chemiluminescence using a platinum electrode. The following conditions were suitable for the determination of glutathione: running buffer, 0.05 mol L−1 Na2HPO4–NaH2PO4 (pH 8.0); concentration of tris(2,2′-bipyridine) ruthenium(II) (Ru(bpy)32+), 5.00 × 10−3 mol L−1; injection voltage, 15 kV; injection time 13 s; separation voltage, 15 kV; detection potential, 1.2 V (vs. Ag/AgCl). Individual Ramos cells were injected into a fused-silica capillary by electromigration and lysed by 0.1% sodium dodecylsulfate. The concentration limit of detection of the method (S/N = 3) was 6.2 × 10−7 mol L−1. The linear dynamic range was from 1.58 × 10−6 to 2.00 × 10−4 mol L−1 with correlation coefficient of 0.9967. The recovery was in the range of 97–103%, and the amount of glutathione in single Ramos cells ranged from 22.0–28.0 fmol. This method indicates a good reproducibility for single cell analysis to provide accurate analytical results.
No takes yet. Share an insight, caveat, or question.
Ji et al. (2013) studied this question.