Key result
MRI-measured ablation gap lengths correlate strongly with gross pathology in a preclinical model.
Why the study?
Can conduction block occur across ablation line gaps, and can these gaps be visualized acutely using MRI?
Can conduction block occur across ablation line gaps, and can these gaps be visualized acutely using MRI?
Ablation line gaps causing acute conduction block can potentially allow arrhythmia recurrences over time, but these gaps can be visualized acutely using MRI to guide targeted ablation.
Answers are grounded in this paper and cite their sources.
MRI gap measurements correlate with pathology in canines; leaves open clinical utility for guiding ablation.
BACKGROUND: Ablation has become an important tool in treating atrial fibrillation and ventricular tachycardia, yet the recurrence rates remain high. It is well established that ablation lines can be discontinuous and that conduction through the gaps in ablation lines can be affected by tissue heating. In this study, we looked at the effect of tissue conductivity and propagation of electric wave fronts across ablation lines with gaps, using both simulations and an animal model. METHODS AND RESULTS: For the simulations, we implemented a 2-dimensional bidomain model of the cardiac syncytium, simulating ablation lines with gaps of varying lengths, conductivity, and orientation. For the animal model, transmural ablation lines with a gap were created in 7 mongrel dogs. The gap length was progressively decreased until there was conduction block. The ablation line with a gap was then imaged using MRI and was correlated with histology. With normal conductivity in the gap and the ablation line oriented parallel to the fiber direction, the simulation predicted that the maximum gap length that exhibited conduction block was 1.4 mm. As the conductivity was decreased, the maximum gap length with conduction block increased substantially, that is, with a conductivity of 67% of normal, the maximum gap length with conduction block increased to 4 mm. In the canine studies, the maximum gap length that displayed conduction block acutely as measured by gross pathology correlated well (R(2) of 0.81) with that measured by MRI. CONCLUSIONS: Conduction block can occur across discontinuous ablation lines. Moreover, with recovery of conductivity over time, ablation lines with large gaps exhibiting acute conduction block may recover propagation in the gap over time, allowing recurrences of arrhythmias. The ability to see gaps acutely using MRI will allow for targeting these sites for ablation.
No takes yet. Share an insight, caveat, or question.
Ranjan et al. (2011) studied Arrhythmia (preclinical ablation model) (n=7). Transmural ablation lines with a gap vs. Pre-ablation baseline / no gap was evaluated on Correlation of gap length displaying conduction block measured by gross pathology versus MRI (R2 = 0.81). In a canine model, the maximum ablation gap length displaying conduction block acutely as measured by gross pathology correlated well (R2=0.81) with gap length measured by MRI.