Key result
The 8th order Butterworth digital filter significantly improved ECG signal quality compared to the analog filter across all leads (p<0.05), displaying smoother signals with less noise.
Why the study?
Obtaining high-quality ECG signals with portable remote acquisition systems is challenging, motivating comparison of analogue and digital filtering for easier signal reading.
Does an 8th order Butterworth digital filter improve ECG signal smoothness compared to a 2nd order analogue filter?
Does an 8th order Butterworth digital filter improve ECG signal smoothness compared to a 2nd order analogue filter?
An 8th order Butterworth digital filter provides smoother ECG signals compared to a standard analogue filter, potentially improving remote ECG analysis.
Answers are grounded in this paper and cite their sources.
May support digital filtering for smoother ECG signals in devices; leaves open clinical diagnostic impact.
The electrocardiogram (ECG) of the human body is an important basis in heart function as well as the diagnosis of cardiovascular diseases, which has a very vital role in clinical diagnosis. Obtaining high-quality ECG signals with a portable remote ECG acquisition system is a big challenge given limited resources. According to the World Health Organization (WHO), disorders of the cardiovascular system still rank high, causing about 31% of deaths globally. This is because the symptoms of cardiovascular disease cannot be seen directly, but rather by conducting an electrocardiograph (ECG) examination. The purpose of this research is to record the electrical signals of the heart by comparing the 2nd order AD8232 module analogue filter with the 8th order Butterworth digital filter which is a better signal for easier reading. This research uses a multiplexer circuit for switching leads, AD8232 ECG module, 50Hz notch filter circuit, non-inverting amplifier, adder, Arduino Mega 2560, USB module, and an application to display digital signals, namely Delphi 7. Signal acquisition is done by monitoring for one minute. Data collection was carried out with 5 respondents 5 times on each lead. The results of the data collection can be concluded that 80% of digital filters display smoother signals for ECG signals than analogue filters
No takes yet. Share an insight, caveat, or question.
Saputri et al. (2022) studied Healthy (ECG signal quality assessment) (n=5). 8th order Butterworth digital filter (IIR) vs. 2nd order AD8232 module analogue filter was evaluated on ECG signal error percentage (p=<0.05). The 8th order Butterworth digital filter significantly improved ECG signal quality compared to the analog filter across all leads (p<0.05), displaying smoother signals with less noise.
Synapse has enriched 10 related papers. Consider them for comparative context: