Key result
Modified PPVI without prestenting cuts procedure duration by ~35 minutes vs. conventional technique.
Why the study?
A modified SAPIEN 3 PPVI technique without prestenting using a protective valve delivery method was developed, prompting comparison of its procedural results with conventional technique.
Does a modified percutaneous pulmonary valve implantation technique without systematic prestenting reduce procedure time and complications in patients with RVOT dysfunction?
Cohort (n=54)
No
Does a modified percutaneous pulmonary valve implantation technique without systematic prestenting reduce procedure time and complications in patients with RVOT dysfunction?
A modified percutaneous pulmonary valve implantation technique without prestenting utilizing a protective delivery method safely reduces procedural time and fluoroscopy exposure while maintaining good hemodynamic outcomes.
Introduction: Percutaneous pulmonary valve implantation (PPVI) with a SAPIEN 3 valve is effective for treating treat right ventricle outflow (RVOT) dysfunction. A modified technique was developed without prestenting using a protective valve delivery method. We aimed to compare the procedural results of the modified technique group (MTG) to those of patients in a conventional technique group (CTG). Methods: We designed a matched before–after study. All consecutive PPVI with SAPIEN 3 performed in the MTG over 9 months were matched, based on the RVOT type and size, to consecutive procedures performed previously with SAPIEN 3. Results: A total of 54 patients were included, equally distributed in the two groups. The sizes of the SAPIEN 3 valves were 23 mm (n = 9), 26 mm (n = 9), 29 mm (n = 36). The two groups were similar regarding demographic data, RVOT type, and pre-procedure hemodynamics. PPVI was performed in a single procedure in all patients of the MTG, whereas six (22.2%) patients of the CTG group underwent prestenting as a first step and valve implantation later (p = 0.02). The procedures were successful in all cases. Stent embolization was reported in two patients (7.4%) in the CTG, which were impacted in pulmonary arteries. In one case (3.7%), in the MTG, an unstable 29 mm SAPIEN 3 valve was stabilized with two stents and additional valve-in-valve implantation. The hemodynamics results were good in all cases, without significant differences between the two groups. The procedures’ durations and fluoroscopy times were significantly reduced in the MTG (48.1 versus 82.6 min, p < 0.0001; 15.2 versus 29.8 min, p = 0.0002). During follow-up, neither stent fracture nor valve dysfunction was noticed in either group. Conclusion: PPVI without prestenting and with a protective delivery method of the SAPIEN 3 valve significantly reduces the procedure’s complexity, the duration, and the irradiation while maintaining excellent hemodynamics results in selected cases.
No takes yet. Share an insight, caveat, or question.
Houeijeh et al. (2023) conducted a cohort in Right ventricle outflow tract (RVOT) dysfunction (n=54). Modified percutaneous pulmonary valve implantation (without prestenting, using large GORE DrySeal Flex sheath) vs. Conventional percutaneous pulmonary valve implantation (with prestenting) was evaluated on Procedure duration (minutes) (p=<0.0001). The modified technique for percutaneous pulmonary valve implantation without prestenting significantly reduced procedure duration compared to the conventional technique (48.1 vs 82.6 minutes, p<0.0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: