Key result
Novel bioresorbable magnesium stents demonstrate minimal in-stent hyperplasia with ~3% MACE at six months.
Why the study?
Permanent metallic DES cause blooming artifacts on CCTA, whereas a novel sirolimus-eluting bioresorbable magnesium-based stent may overcome vessel caging without the thrombosis risks of earlier scaffolds.
Does a novel bioresorbable magnesium-based sirolimus-eluting stent maintain lumen patency and allow for artifact-free CCTA evaluation at 6 months in patients with de novo coronary lesions?
Observational (n=29)
No
Does a novel bioresorbable magnesium-based sirolimus-eluting stent maintain lumen patency and allow for artifact-free CCTA evaluation at 6 months in patients with de novo coronary lesions?
Absolute Event Rate: 6.3% vs 7.3%
p-value: p=0.155
A novel bioresorbable magnesium-based sirolimus-eluting stent demonstrated minimal in-stent hyperplasia at 6 months and allowed for fully interpretable, artifact-free non-invasive evaluation by CCTA.
Supports feasibility of magnesium stents with artifact-free CCTA; leaves open confirmation of safety and efficacy in randomized trials.
Background Drug-eluting stents (DES) are the gold standard in percutaneous coronary interventions (PCI), but leave a permanent metallic “caging” of the treated vessels limiting further assessment of in-stent lumen patency by coronary computed tomography angiography (CCTA) due to artifacts. Absorbable scaffolds were designed to overcome the caging limitation but the first generation made of poly-L-lactic acid demonstrated higher thrombosis rates. A novel bioresorbable magnesium-based (Mg-)stent coated with a biodegradable polymer eluting sirolimus has been developed, with promising results in the BIOSOLVEII-III studies. Aim We sought to characterize PCI results by CCTA at 6-mo follow-up of patients treated with at least one Mg-stent in our institution. Methods Prospective observational registry started since January 2017 of younger patients with de novo lesions preferably treated with Mg-stents after balloon pre-dilatation. Procedural data and major adverse clinical events (MACE) at hospital discharge and 6-mo follow-up were collected. Reference vessel and in-stent minimal lumen area were measured on a CCTA performed at 6-mo. Results 34 Mg-stents (mean diameter: 3.2±0.2 mm, length 21.3±4.1 mm) were successfully implanted in 29 patients (mean age 54±6 years with male:female ratio 3:1). Acute coronary syndrome was the presenting diagnosis in 76% (n=22) with STEMI in 31% (n=9). The left anterior descending artery was treated in 62% (n=21). Calcifications on angiography were found in 14 lesions (41%). Intravascular imaging was performed in 3 PCI. With CCTA at 7.1±3.5 months (n=15 up to date, ongoing further follow-up to be presented), proximal and distal stent markers were well visualized while scaffold struts were not discernible. Mean proximal and distal reference lumen area were respectively 8.5±4.2 mm2 and 6.1±2.8 mm2. Mean in-stent minimal lumen area (MLA) was 6.3±2.9 mm2, with no statistical difference with the mean of the proximal and distal references (7.3±3.3 mm2, p=0.155, Wilcoxon rank test) demonstrating minimal instent hyperplasia at 6-mo: significant in-stent restenosis was noted in only one patient who remains so far asymptomatic (MLA 1.1 mm2; reference vessel lumen area 5.5 mm2). CCTA were non interpretable in 2 patients due to artifacts unrelated to the Mg-stents. One death secondary to a complicated cardiac tamponade was reported. No further MACE at 6-mo were noted. CCTA cross sectional cut Conclusion 6-mo CCTA of patients treated with a Mg-stent are fully interpretable to detect in-stent restenosis, without blooming artifacts. Accurate non-invasive assessment of the late results of our monocentric observational registry demonstrated 1 asymptomatic instent restenosis in 34 Mg-stents (3%) and overall optimal stent deployment and late artery patency, achieved with only 1 MACE in 29 patients (3.4%). This highlights the potentials of this new Mg-bioresorbable stent and the use of CCTA for clinical follow-up of the treated patients.
No takes yet. Share an insight, caveat, or question.
Ghafari et al. (2019) conducted an observational in de novo coronary lesions (n=29). Bioresorbable magnesium-based (Mg-)stent vs. Proximal and distal reference lumen area was evaluated on Mean in-stent minimal lumen area (MLA) at 6 months (p=0.155). A novel bioresorbable magnesium-based stent demonstrated minimal in-stent hyperplasia at 6 months, with a mean minimal lumen area of 6.3 mm2, 3% in-stent restenosis, and 3.4% MACE.