Key result
Visualized steerable sheath linked to ~1.4 fewer minutes of transseptal fluoroscopy vs. conventional sheath.
Observational (n=60)
Reveals improved outcomes in catheter stability and fluoroscopy times during pulmonary vein isolation.
BACKGROUND: A novel steerable sheath visualized on a three-dimensional mapping system has become available in this era in which a durable pulmonary vein (PV) isolation (PVI) with reduced fluoroscopy is required. METHODS: In 60 patients who underwent a PVI with a visualized sheath (n = 30) and non-visualized conventional sheath (n = 30), the fluoroscopic time and catheter stability during the PVI were analyzed. RESULTS: The fluoroscopic time during the transseptal access (0 [0, 0.1] vs. 1.4 [0.8, 2.3] minutes, P < .001) and PVI (0 [0, 0.1] vs. 0.4 [0.2, 1.1] minutes, P < .001) were shorter in the visualized sheath group than conventional sheath group. The procedure time during the PVI (32.0 [26.8, 36.3] vs. 41.0 [31.8, 47.3] minutes, P = .01), particularly during the right PVI (15.0 [12.8, 18.0] vs. 23.0 [15.8, 26.3] minutes, P = .009), was shorter in the visualized sheath group than conventional sheath group, however, that during the other steps was equivalent. The standard deviation of the catheter contact force during each radiofrequency application was smaller in the visualized sheath group than conventional sheath group (4.5 ± 2.7 vs. 4.9 ± 3.1 g, P = .001). The impedance drop for each lesion was larger in the visualized sheath group than conventional sheath group (10.7 ± 6.5 vs. 9.8 ± 5.5 ohms, P < .001). The incidence of acute PV reconnections per patient (30% vs. 23%, P = .56) and per PV segment (2.5% vs. 2.3%, P = .83) were similar between the two groups. No major complications occurred in either sheath group. CONCLUSIONS: The use of visualized sheaths may reduce the fluoroscopic time and improve the catheter stability during the PVI.
No takes yet. Share an insight, caveat, or question.
Wakamatsu et al. (2021) conducted an observational in Atrial fibrillation (undergoing pulmonary vein isolation) (n=60). Steerable sheath visualized by a 3D mapping system and intracardiac ultrasound vs. Non-visualized conventional sheath was evaluated on Fluoroscopic time during transseptal access (minutes) (p=<0.001). The use of a visualized steerable sheath significantly reduced the median fluoroscopic time during transseptal access compared to a conventional sheath (0 vs. 1.4 minutes, P<0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: