Key result
In an in vitro model, the discharge coefficient of orifices ranged from 0.675 to 0.93, decreasing by 5.5% with greater eccentricity and increasing by 8.9% with larger area and 8.8% with a nozzle inlet.
Why the study?
Fluid dynamics suggests orifice geometry influences discharge properties and empiric constants in formulas like the Gorlin formula used to calculate stenotic valve area.
In vitro orifice gains should not change practice; hypothesis-generating for prosthetic valve design.
Fluid dynamics suggests that orifice geometry is a determinant of discharge properties and, therefore, should influence empiric constants in formulas (such as the Gorlin formula) to calculate stenotic valve area. An in vitro study utilizing a model of transmitral flow was conducted to investigate how the discharge coefficient changes with 1) orifice eccentricity (ratio of long to short diameter), 2) absolute area, 3) the presence of a nozzle-like inlet, and 4) varying flow. Twenty-three orifices with areas varying between 0.3 and 2.5 cm2 and eccentricities from 1:1, or circular, to 5:1, or elliptic, were tested. The calculated discharge coefficients ranged between 0.675 and 0.93. For a given area, the discharge coefficient decreased by a mean value (+/- SD) of 5.5 +/- 1.3% between circular orifices and 5:1 ellipses. Discharge coefficients increased by a mean of 8.9 +/- 3.5% from 0.3 to 2.5 cm2 area within each eccentricity class. A gradually tapering inlet (nozzle) raised the discharge coefficient by 8.8 +/- 3.9%, leading to a discharge coefficient between 0.81 and 0.93 for round orifices. The discharge coefficient did not change appreciably with flow. The concept of the discharge coefficient and its role in assessing restrictive orifices in general by hydraulic formulas (for example, the Gorlin and pressure half-time calculations) are discussed.
No takes yet. Share an insight, caveat, or question.
Flachskampf et al. (1990) studied Stenotic valve area (n=23). Orifice geometry and flow rate was evaluated on Discharge coefficient. In an in vitro model, the discharge coefficient of orifices ranged from 0.675 to 0.93, decreasing by 5.5% with greater eccentricity and increasing by 8.9% with larger area and 8.8% with a nozzle inlet.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: