BACKGROUND: We performed a multicenter evaluation of the Elecsys® progastrin-releasing peptide (ProGRP) immunoassay in Europe and China. METHODS: The assay was evaluated at three European and two Chinese sites by imprecision, stability, method comparison and differentiation potential in lung cancer. RESULTS: Intermediate imprecision across five analyte concentrations ranged from 2.2% to 6.0% coefficient of variation. Good stability for plasma and serum samples was shown for various storage conditions. There was excellent correlation between the Elecsys® and ARCHITECT assays in plasma (slope 1.02, intercept -2.72pg/mL). The Elecsys® assay also showed good correlation between serum and plasma samples (slope 0.93, intercept 2.35pg/mL; correlation coefficient 0.97). ProGRP differentiated small-cell and non-small-cell lung cancer (NSCLC; area under the curve 0.90, 95% CI 0.87-0.93; 78.3% sensitivity, 95% specificity; at 84pg/mL), with no relevant effects of ethnicity, age, gender or smoking. Median ProGRP concentrations were low in benign diseases (38pg/mL), other malignancies (40pg/mL) or NSCLC (39pg/mL), except chronic kidney disease above stage 3 (>100pg/mL). CONCLUSIONS: Increased stability of the Elecsys® ProGRP assay in serum and plasma offers clear benefits over existing assays. This first evaluation of a ProGRP assay in China demonstrated comparable differentiation potential among different ethnicities.