Key result
Simultaneous dual-isotope acquisition with 123I and 99mTc yielded similar results for left ventricular ejection fraction and regional motion compared to single-isotope acquisition using two CZT cameras.
Why the study?
Does simultaneous dual-isotope acquisition with CZT cameras yield similar left ventricular function assessment compared to single-isotope acquisition in a cardiac phantom model?
Does simultaneous dual-isotope acquisition with CZT cameras yield similar left ventricular function assessment compared to single-isotope acquisition in a cardiac phantom model?
p-value: p=NS
Simultaneous dual-isotope acquisition with CZT cameras provides similar left ventricular function assessment compared to single-isotope acquisition in a cardiac phantom model.
Supports dual-isotope CZT feasibility in phantoms; leaves open clinical validation in patients.
The impact of increased energy resolution of cadmium–zinc–telluride (CZT) cameras on the assessment of left ventricular function under dual-isotope conditions ( 99m Tc and 123 I) remains unknown. The Amsterdam-gated dynamic cardiac phantom (AGATE, Vanderwilt techniques, Boxtel, The Netherlands) was successively filled with a solution of 123 I alone, 99m Tc alone, and a mixture of 123 I and 99m Tc. A total of 12 datasets was acquired with each commercially available CZT camera (DNM 530c, GE Healthcare and DSPECT, Biosensors International) using both energy windows ( 99m Tc or 123 I) with ejection fraction set to 33, 45, and 60 %. End-diastolic (EDV) and end-systolic (ESV) volumes, ejection fraction (LVEF), and regional wall motion and thickening (17-segment model) were assessed using Cedars-Sinai QGS Software. Concordance between single- and dual-isotope acquisitions was tested using Lin’s concordance correlation coefficient (CCC) and Bland–Altman plots. There was no significant difference between single- or simultaneous dual-isotope acquisition ( 123 I and 99m Tc) for EDV, ESV, LVEF, or segmental wall motion and thickening. Myocardial volumes using single- ( 123 I, 99m Tc) and dual-isotope (reconstructed using both 123 I and 99m Tc energy windows) acquisitions were, respectively, the following: EDV (mL) 88 ± 27 vs. 89 ± 27 vs. 92 ± 29 vs. 90 ± 26 for DNM 530c ( p = NS) and 82 ± 20 vs. 83 ± 22 vs. 79 ± 19 vs. 77 ± 20 for DSPECT ( p = NS); ESV (mL) 40 ± 1 vs. 41 ± 2 vs. 41 ± 2 vs. 42 ± 1 for DNM 530c ( p = NS) and 37 ± 5 vs. 37 ± 1 vs. 35 ± 3 vs. 34 ± 2 for DSPECT ( p = NS); LVEF (%) 52 ± 14 vs. 51 ± 13 vs. 53 ± 13 vs. 51 ± 13 for DNM 530c ( p = NS) and 52 ± 16 vs. 54 ± 13 vs. 54 ± 14 vs. 54 ± 13 for DSPECT ( p = NS); regional motion (mm) 6.72 ± 2.82 vs. 6.58 ± 2.52 vs. 6.86 ± 2.99 vs. 6.59 ± 2.76 for DNM 530c ( p = NS) and 6.79 ± 3.17 vs. 6.81 ± 2.75 vs. 6.71 ± 2.50 vs. 6.62 ± 2.74 for DSPECT ( p = NS). The type of camera significantly impacted only on ESV ( p < 0.001). The new CZT cameras yielded similar results for the assessment of LVEF and regional motion using different energy windows ( 123 I or 99m Tc) and acquisition types (single vs. dual). With simultaneous dual-isotope acquisitions, the presence of 123 I did not impact on LVEF assessment within the 99m Tc energy window for either CZT camera.
No takes yet. Share an insight, caveat, or question.
Blaire et al. (2016) studied Left ventricular function assessment (phantom). Simultaneous dual-isotope acquisition (123I and 99mTc) vs. Single-isotope acquisition was evaluated on Difference in LVEF, EDV, ESV, and regional wall motion between single- and dual-isotope acquisitions (p=NS). Simultaneous dual-isotope acquisition with 123I and 99mTc yielded similar results for left ventricular ejection fraction and regional motion compared to single-isotope acquisition using two CZT cameras.