PURPOSE: To compare the anterior segment measurements provided by 3 Scheimpflug tomographers and a Placido corneal topographer. SETTING: Private clinical ophthalmology practice. DESIGN: Evaluation of diagnostic test or technology. METHODS: In a sample of 25 consecutive patients having either refractive or cataract surgery, the anterior eye segment was analyzed by means of a rotating Scheimpflug camera (Pentacam), 2 devices with a Scheimpflug camera combined with a Placido disk (Sirius and TMS-5), and a Placido disk corneal topographer (Keratron). Measurement results were compared using analysis of variance. Agreement was assessed using Bland-Altman plots. RESULTS: The mean simulated keratometry (K) was different between the 4 instruments (P<.0001), with Keratron providing the highest value (44.43 diopters [D] ± 1.28 [SD]). The Pentacam and Sirius provided the lowest values (44.05 ± 1.21 D and 44.05 ± 1.27 D, respectively), without statistical difference (posttest). The mean posterior corneal power and minimum corneal thickness were statistically different between the 3 Scheimpflug cameras (P<.0001 and P=.0210, respectively); 95% limits of agreement, however, were narrow for posterior corneal power and large for corneal thickness. The only 2 devices measuring the distance between the corneal endothelium and the anterior lens surface showed a statistically but not clinically significant difference (2.90 ± 0.48 mm and 2.94 ± 0.47 mm, respectively). There were no statistically significant differences in anterior corneal asphericity between the 4 instruments. CONCLUSION: Although the measurements of some parameters by different instruments were similar, caution is warranted before using them interchangeably.
No takes yet. Share an insight, caveat, or question.
Savini et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: