PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
November 1, 1995European Heart Journal97 citations

A comparison of non-invasive continuous finger blood pressure measurement (Finapres) with intra-arterial pressure during prolonged head-up tilt

View Full Paper
MPMARK E.V. PETERSENTWTimothy R. WilliamsRSRichard Sutton

Key Result

Non-invasive continuous finger blood pressure measurement (Finapres) closely followed intra-arterial pressure during head-up tilt, with a mean systolic bias of +0.7 mmHg and diastolic of +5.4 mmHg.

Key Points

  • To compare non-invasive finger blood pressure measurements using Finapres with intra-arterial measurements during prolonged head-up tilt in patients with vasovagal syncope.
  • Eight patients with malignant vasovagal syncope underwent twelve tilt tests.
  • Systolic and diastolic pressures were measured simultaneously by both non-invasive and intra-arterial methods.
  • Data on mean bias and within-tilt precision were analyzed.
  • Mean bias for systolic pressure was +0.7 mmHg with a standard deviation of 11.3 mmHg.
  • For diastolic pressure, mean bias was +5.4 mmHg with a standard deviation of 7 mmHg.
  • Non-invasive measurements tracked 89% of beat-to-beat systolic changes and 95% of diastolic changes within +/- 2 mmHg.

Study Design

Type

Observational (n=8)

Structured PICO

Does non-invasive continuous finger blood pressure measurement (Finapres) accurately reflect intra-arterial pressure during prolonged head-up tilt in patients with malignant vasovagal syncope?

P
Population
8 patients (mean age 49 years) with malignant vasovagal syncope undergoing prolonged head-up tilt testing.
E
Exposure
Non-invasive continuous finger blood pressure measurement (Finapres) during prolonged head-up tilt.
C
Comparator
Simultaneous intra-radial (intra-arterial) blood pressure measurement.
O
Outcome
Mean bias (difference between Finapres and intra-arterial pressures) for systolic and diastolic pressure.surrogate

Non-invasive continuous finger blood pressure measurement (Finapres) demonstrates acceptable precision and closely tracks intra-arterial pressure changes during diagnostic tilt testing.

Main Result

Mean Difference: 0.7

Abstract

Simultaneous intra-radial and non-invasive (Finapres, Ohmeda) blood pressures were compared during prolonged head-up tilt, in eight patients (mean age 49 years) with malignant vasovagal syncope. Twelve tilts were performed, of which eight resulted in vasovagal syncope. The mean bias (difference between Finapres and intra-arterial pressures) for systolic pressure was +0.7 mmHg (standard deviation 11.3 mmHg) and for diastolic pressure was +5.4 mmHg (standard deviation 7 mmHg). The within-tilt precision (standard deviation of the bias) of the non-invasive measurements varied between 2.9-12.4 mmHg (median 4.5 mmHg) for systolic comparisons, and 1.6-8.4 mmHg (median 4.4 mmHg) for diastolic comparisons. In all but one tilt highly significant positive increases in both systolic (median 7.1 mmHg) and diastolic bias (median 8.1 mmHg) occurred on tilt with respect to resting pre-tilt levels. Independent of the absolute level of agreement, the non-invasive measurements followed changes in intra-arterial pressure closely, with 89% of beat-to-beat changes in systolic pressure, and 95% of beat-to-beat changes in diastolic pressure followed to within +/- 2 mmHg. This study suggests that the Finapres is well suited for use during diagnostic tilt testing, demonstrating an acceptable within-tilt precision and closely following pressure changes during vasovagal syncope.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

PETERSEN et al. (1995) conducted an observational in Malignant vasovagal syncope (n=8). Non-invasive continuous finger blood pressure measurement (Finapres) vs. Intra-arterial pressure measurement was evaluated on Mean bias (difference between Finapres and intra-arterial pressures) for systolic and diastolic pressure (Mean bias +0.7 mmHg (systolic) and +5.4 mmHg (diastolic)). Non-invasive continuous finger blood pressure measurement (Finapres) closely followed intra-arterial pressure during head-up tilt, with a mean systolic bias of +0.7 mmHg and diastolic of +5.4 mmHg.

synapsesocial.com/papers/6a6134bb8f214046fa3f2e3chttps://doi.org/10.1093/oxfordjournals.eurheartj.a060791
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Finger Arterial versus Intrabrachial Pressure and Continuous Cardiac output during Head-up Tilt Testing in Healthy Subjects1996 · 121 citations
  2. 2Continuous Noninvasive Finger Blood Pressure during Controlled Hypotension A Comparison with Intraarterial Pressure1991 · 29 citations
  3. 3Continuous non-invasive blood pressure monitoring: reliability of Finapres device during the Valsalva manoeuvre1988 · 276 citations
  4. 4The Reliability of Noninvasive Continuous Finger Blood Pressure Measurement in Patients With Both Hypertension and Vascular Disease1992 · 85 citations
  5. 5Comparison of intrabrachial and finger blood pressure in healthy elderly volunteers1995 · 84 citations