Validates neutron emission measurements in the KATANA facility, supporting future ITER operations.
KATANA is a water activation facility located in the TRIGA Mark II reactor of the Jožef Stefan Institute in Ljubljana, Slovenia. Since its commissioning it has contributed to a better understanding of water activation processes and their modelling, which is important in the context of the future operation of the ITER tokamak. The facility is designed to provide a flexible, benchmark-quality environment for the validation of computational tools for the activation of water. It will also support shielding experiments, in particular for ITER-relevant materials, and serve as a stable source of high-energy gamma rays and neutrons. This paper presents the calibration of 3 He and BF 3 neutron detectors used to measure neutron emission from the decay of 17 N. The calibration was performed at the Czech Metrology Institute using a well-characterised graphite-moderated thermal neutron reference field. The calibrated acquisition system is then used in conjunction with Monte Carlo particle transport simulations to obtain the total neutron emission within the measurement volume of the KATANA water activation facility. The results form a basis for the determination of the integral cross section of the 17 O(n,p) 17 N reaction, a nuclear reaction that is still insufficiently described in the literature.