Techniques to image the microvasculature may play an important role in imaging tumor-related angiogenesis and vasa vasorum associated with vulnerable atherosclerotic plaques. However, the microvasculature associated with these pathologies is difficult to detect using traditional B-mode ultrasound or even harmonic imaging due to small vessel size and poor differentiation from surrounding tissue. Acoustic angiography, a microvascular imaging technique that utilizes superharmonic imaging (detection of higher order harmonics of microbubble response), can yield a much higher contrast-to-tissue ratio than second harmonic imaging methods. In this paper, two dual-frequency transducers using lateral mode transmitters were developed for superharmonic detection and acoustic angiography imaging in intracavity applications. A single element dual-frequency intravascular ultrasound transducer was developed for concept validation, which achieved larger signal amplitude, better contrast-to-noise ratio (CNR), and pulselength compared to the previous work. A dual-frequency [Pb(M g1/3 Nb 2/3 )O 3 ]-x[PbTiO 3 ] array transducer was then developed for superharmonic imaging with dynamic focusing. The axial and lateral sizes of the microbubbles in a 200-μm tube were measured to be 269 and 200 μm, respectively. The maximum CNR was calculated to be 22 dB. These results show that superharmonic imaging with a low frequency lateral mode transmitter is a feasible alternative to thickness mode transmitters when the final transducer size requirements dictate design choices.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: