Key result
A newly developed 10-Fr 4-D ICE catheter provided 2-D image quality comparable to commercial catheters and a 4-D field of view of at least 90° × 60° × 8 cm at 30 vol/s in preclinical studies.
Answers are grounded in this paper and cite their sources.
A newly developed 4-D ICE catheter provides real-time 3-D imaging with a wide field of view, potentially reducing fluoroscopy use in EP procedures and replacing TEE in structural heart interventions.
We developed a 2.5×6.6 mm 2 2-D array transducer with integrated transmit/receive application-specific integrated circuit (ASIC) for real-time 3-D intracardiac echocardiography (4-D ICE) applications. The ASIC and transducer design were optimized so that the high-voltage transmit, low-voltage time-gain control and preamp, subaperture beamformer, and digital control circuits for each transducer element all fit within the 0.019-mm 2 area of the element. The transducer assembly was deployed in a 10-Fr (3.3-mm diameter) catheter, integrated with a GE Vivid E9 ultrasound imaging system, and evaluated in three preclinical studies. The 2-D image quality and imaging modes were comparable to commercial 2-D ICE catheters. The 4-D field of view was at least 90° × 60° × 8 cm and could be imaged at 30 vol/s, sufficient to visualize cardiac anatomy and other diagnostic and therapy catheters. 4-D ICE should significantly reduce X-ray fluoroscopy use and dose during electrophysiology ablation procedures. 4-D ICE may be able to replace transesophageal echocardiography (TEE), and the associated risks and costs of general anesthesia, for guidance of some structural heart procedures.
No takes yet. Share an insight, caveat, or question.
Wildes et al. (2016) studied this question. 4-D ICE catheter (2-D array transducer with integrated ASIC) vs. Commercial 2-D ICE catheters was evaluated on 2-D image quality, imaging modes, and 4-D field of view. A newly developed 10-Fr 4-D ICE catheter provided 2-D image quality comparable to commercial catheters and a 4-D field of view of at least 90° × 60° × 8 cm at 30 vol/s in preclinical studies.
Synapse has enriched 10 related papers. Consider them for comparative context: