Arterial oxygen tension measurements were performed simultaneously using two different techniques: (1) the conventional method of analyzing a blood sample obtained from the radial artery by means of a Clark electrode and (2) a new method of transcutaneous oxygen tension recording using a newly developed surface electrode containing a built-in heating device to ensure optimal cutaneous perfusion at the site of measurement. Two groups of newborn infants were used as subjects: (1) 70 clinically healthy babies who were tested during normoxia and hyperoxia (breathing 80% to 100% oxygen) and (2) 20 sick preterm and term infants receiving inspired oxygen concentrations of between 21% and 100% during the measurement. Our results indicate a satisfactory accuracy for the transcutaneous oxygen tension measurements in normoxia and hyperoxia (percentage coefficient of variation, 15.9% and 24.1%, respectively). In hypoxia agreement between the two methods varies depending on the degree of circulatory derangement. Overall correlation coefficients were greater than 0.85 in each group.
No takes yet. Share an insight, caveat, or question.
Fenner et al. (1975) studied this question.