Key result
Electrogram analysis during left ventricular pacing tests reliably detected cathodal or anodal capture (98% morphological consistency), offering greater sensitivity than 12-lead ECG changes.
Why the study?
The study aimed to identify cardiac electrogram characteristics for detecting cathodal, anodal, or cathodal-anodal capture in left ventricular quadripolar pacing leads of CRT patients, and examine their relationship with the ECG.
Does EGM analysis reliably detect cathodal, anodal, or cathodal-anodal capture compared to 12 lead ECG in CRT patients with quadripolar leads?
Observational (n=9)
Does EGM analysis reliably detect cathodal, anodal, or cathodal-anodal capture compared to 12 lead ECG in CRT patients with quadripolar leads?
EGM characteristics during LV pacing tests can reliably detect cathodal, anodal, or cathodal-anodal capture with greater sensitivity than 12-lead ECG changes, potentially improving routine CRT device follow-up.
EGM analysis may refine LV capture detection in CRT; hypothesis-generating and requires prospective validation before practice change.
AIMS: Our study analyzed cardiac electrograms (EGMs) to identify characteristics for detecting cathodal, anodal, or cathodal-anodal (simultaneous) capture in left ventricular (LV) quadripolar pacing leads of cardiac resynchronization therapy (CRT) patients. The relationship between these EGM characteristics and the electrocardiogram (ECG) was also examined. METHODS: We performed a retrospective analysis of 54 bipolar pacing configurations across nine patients with implanted CRT devices and quadripolar leads who had undergone a 12 lead ECG optimization. Three pacing tests (cathode unipolar, anode unipolar, and bipolar) per bipolar pair were performed, examining ECG and EGM morphology changes accompanying each test and any transitions of morphology or amplitude during voltage stepdown. RESULTS: During the cathode and anode unipolar pacing tests, the EGM was biphasic (negative/positive) or monophasic (positive) in 52/53 (98%), and biphasic (positive/negative) or monophasic (negative) in 50/51 (98%), respectively. During bipolar LV capture threshold testing, 30 bipolar pairs displayed a sudden increase in EGM amplitude (median 9.4 mV, interquartile range [7-14 mV]) when transitioning from cathodal-anodal capture to cathodal or anodal capture. Ninety percent of these EGM transitions had a corresponding simultaneous change in ECG, while 10% had no ECG changes. Two patients demonstrated "quad-site" capture on their quadripolar lead with multipoint pacing enabled and cathodal-anodal capture from each stimulus. CONCLUSION: EGM characteristics during LV pacing tests can reliably detect cathodal, anodal, or cathodal-anodal capture, with greater sensitivity than 12 lead ECG changes. Integration of EGM analysis into routine CRT device follow up can be performed easily and may have implications for CRT efficacy.
No takes yet. Share an insight, caveat, or question.
Pfeffer et al. (2021) conducted an observational in Cardiac resynchronization therapy (CRT) (n=9). Electrogram (EGM) analysis vs. 12-lead ECG was evaluated on Detection of cathodal, anodal, or cathodal-anodal capture. Electrogram analysis during left ventricular pacing tests reliably detected cathodal or anodal capture (98% morphological consistency), offering greater sensitivity than 12-lead ECG changes.