January 30, 2014AJP Gastrointestinal and Liver Physiology

Water-perfused esophageal high-resolution manometry: normal values and validation

View Full Paper
Ask AI
Bookmark
Share

Key result

Water-perfused high-resolution manometry showed moderate to good agreement with solid-state HRM (e.g., ICC 0.90 for distal contractile integral), with only small differences in outcome measures.

Why the study?

Does water-perfused HRM show acceptable agreement with solid-state HRM in healthy volunteers?

Population

50 healthy volunteers (mean age 35 yr, range 21-64 yr; 15 women, 35 men)

Comparison

Water-perfused high-resolution manometry (HRM) vs Solid-state HRM

Design

Cross-sectional, randomized order of tests

Authors

BKBoudewijn F. KessingUniversity Medical CenterPWPim W. WeijenborgAmsterdam UMC Location University of AmsterdamASA. J. P. M. SmoutAmsterdam University Medical Centers

Discussion

Loading...

Member takes

Implication

May support water-perfused HRM adoption; leaves open validation in symptomatic patients.

Study Design

Type

RCT (n=50)

Randomization

Randomized order (crossover)

Structured PICO

Does water-perfused HRM show acceptable agreement with solid-state HRM in healthy volunteers?

P
Population
50 healthy volunteers (mean age 35 years, 30% female) who underwent both water-perfused and solid-state high-resolution manometry in a randomized order.
I
Intervention
Water-perfused high-resolution manometry (HRM)
C
Comparator
Solid-state HRM
O
Outcome
Normal values (5th-95th percentile ranges) and agreement (intraclass correlation coefficient) between water-perfused and solid-state HRM for contractile front velocity (CFV), distal contractile integral (DCI), distal contraction latency (DL), and integrated relaxation pressure (IRP 4s)surrogate

Main Result

Effect estimate: ICC 0.39 to 0.90

Water-perfused HRM yields normal values comparable to solid-state HRM with moderate to good agreement, validating its use for esophageal manometry.

Cite This Study

Kessing et al. (2014) conducted an RCT in Healthy volunteers (n=50). Water-perfused high-resolution manometry (HRM) vs. Solid-state HRM was evaluated on Agreement between water-perfused and solid-state HRM (assessed via intraclass correlation coefficient) (ICC 0.39 to 0.90). Water-perfused high-resolution manometry showed moderate to good agreement with solid-state HRM (e.g., ICC 0.90 for distal contractile integral), with only small differences in outcome measures.

synapsesocial.com/papers/6a6344e8cd7899abd6b6fbc1https://doi.org/10.1152/ajpgi.00447.2013
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Normal values for esophageal high‐resolution manometry2013 · 89 citations
  2. 2Chicago classification criteria of esophageal motility disorders defined in high resolution esophageal pressure topography12012 · 778 citations
  3. 3Characterization of the pharyngo-UES contractile reflex in humans1997 · 68 citations
  4. 4Quantifying EGJ morphology and relaxation with high-resolution manometry: a study of 75 asymptomatic volunteers2006 · 218 citations