The objective imaging characteristics of three systems that use charge coupled devices (CCD) for small-field digital mammography (SFDM) have been compared in terms of spatial resolution and signal to noise ratio. The results indicate that although they are designed for nominally the same tasks of stereotactic localization and spot imaging these detectors have significantly differing physical imaging properties. Imaging system design parameters such as the phosphor screen type and thickness, screen configuration and method of optically coupling the phosphor to the CCD have significant effects on the imaging performance of the detectors.