Key result
Biplanar imaging methods showed good agreement with 3D techniques for assessing stent geometry, with a maximum disagreement of +/- 0.03 mm in length and +/- 3 degrees in angle.
Biplanar imaging methods show good agreement with 3D techniques for quantifying stent geometry and deformation, offering advantages in measuring dynamic 3D deformation due to shorter acquisition times.
May facilitate dynamic stent deformation studies in preclinical models; leaves open clinical validation versus 3D techniques.
PURPOSE: To quantify variability of in vitro and in vivo measurement of 3D device geometry using 3D and biplanar imaging. METHODS: Comparison of stent reconstruction is reported for in vitro coronary stent deployment (using micro-CT and optical stereo-photogrammetry) and in vivo pulmonary valve stent deformation (using 4DCT and biplanar fluoroscopy). Coronary stent strut length and inter-strut angle were compared in the fully deployed configuration. Local (inter-strut angle) and global (dog-boning ratio) measures of stent deformation were reported during stent deployment. Pulmonary valve stent geometry was assessed throughout the cardiac cycle by reconstruction of stent geometry and measurement of stent diameter. RESULTS: Good agreement was obtained between methods for assessment of coronary stent geometry with maximum disagreement of +/- 0.03 mm (length) and +/- 3 degrees (angle). The stent underwent large, non-uniform, local deformations during balloon inflation, which did not always correlate with changes in stent diameter. Three-dimensional reconstruction of the pulmonary valve stent was feasible for all frames of the fluoroscopy and for 4DCT images, with good correlation between the diameters calculated from the two methods. The largest compression of the stent during the cardiac cycle was 6.98% measured from fluoroscopy and 7.92% from 4DCT, both in the most distal ring. CONCLUSIONS: Quantitative assessment of stent geometry reconstructed from biplanar imaging methods in vitro and in vivo has shown good agreement with geometry reconstructed from 3D techniques. As a result of their short image acquisition time, biplanar methods may have significant advantages in the measurement of dynamic 3D stent deformation.
No takes yet. Share an insight, caveat, or question.
Zwierzak et al. (2014) studied Stent deployment. Biplanar imaging (optical stereo-photogrammetry and biplanar fluoroscopy) vs. 3D imaging (micro-CT and 4DCT) was evaluated on Stent geometry and deformation (strut length, inter-strut angle, diameter). Biplanar imaging methods showed good agreement with 3D techniques for assessing stent geometry, with a maximum disagreement of +/- 0.03 mm in length and +/- 3 degrees in angle.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: