Key result
Redundant suture knot causes acute severe aortic regurgitation by preventing tilting disc prosthesis occlusion.
Why the study?
Acute valvular dysfunction immediately following SAVR is a dramatic intraoperative complication demanding rapid diagnosis, prompting the report of a mechanical obstruction mimicking refractory myocardial instability.
Case Report (n=1)
No
Early intraoperative TEE is crucial for differentiating mechanical surgical complications from primary myocardial issues in patients with refractory instability after aortic valve replacement.
Answers are grounded in this paper and cite their sources.
Knot-related prosthetic obstruction may mimic refractory instability post-SAVR; this Level 5 case leaves open its true incidence and prevention.
The occurrence of acute valvular dysfunction immediately following surgical aortic valve replacement (SAVR) is a dramatic intraoperative complication that demands rapid diagnosis. [1,2]e wish to share a case of "a naughty knot" -a simple mechanical obstruction that mimicked refractory myocardial instability.A patient with a bicuspid aortic valve and severe calcific aortic stenosis, featuring preserved left ventricular function, underwent elective SAVR.A 25-mm TTK Chitra tilting disc prosthesis was implanted using standard techniques.The cardiopulmonary bypass (CPB) time was 168 min, with an aortic cross-clamp time of 78 min.Following a standard de-airing protocol and clamp removal, the patient repeatedly experienced ventricular fibrillation (VF) during attempts to wean from CPB.Despite correction of electrolytes and four episodes of asynchronous cardioversion, the arrhythmia persisted.CPB was reinstated, the cross-clamp reapplied, and "hot shot" cardioplegia administered.Upon a second attempt to wean, the patient required maximal inotropic support.
No takes yet. Share an insight, caveat, or question.
Yadav et al. (2026) conducted a case report in Severe calcific aortic stenosis with bicuspid aortic valve (n=1). Surgical aortic valve replacement with a 25-mm TTK Chitra tilting disc prosthesis was evaluated on Acute valvular dysfunction and ventricular fibrillation. A redundant suture knot caused acute severe aortic regurgitation by preventing complete occlusion of the tilting disc prosthesis, which was successfully resolved by trimming the extra suture.
Synapse has enriched 10 related papers. Consider them for comparative context: