Key result
During cesarean delivery, a colorimetric system measured blood loss with significantly greater accuracy (mean bias 102 mL) compared to visual estimation (mean bias 458 mL) and gravimetric measurement (mean bias 352 mL).
Why the study?
Does a colorimetric system improve the accuracy of blood loss measurement compared to visual estimation and gravimetric methods in patients having cesarean deliveries?
Observational (n=50)
Clinicians blinded to other assays
No
Does a colorimetric system improve the accuracy of blood loss measurement compared to visual estimation and gravimetric methods in patients having cesarean deliveries?
Colorimetric image analysis is superior to visual estimation and gravimetric methods for measuring blood loss during cesarean delivery.
Answers are grounded in this paper and cite their sources.
Colorimetric estimation may outperform visual and gravimetric methods in cesarean delivery; leaves open routine adoption pending validation.
Objective This study aims to compare the accuracy of visual, quantitative gravimetric, and colorimetric methods used to determine blood loss during cesarean delivery procedures employing a hemoglobin extraction assay as the reference standard. Study Design In 50 patients having cesarean deliveries blood loss determined by assays of hemoglobin content on surgical sponges and in suction canisters was compared with obstetricians' visual estimates, a quantitative gravimetric method, and the blood loss determined by a novel colorimetric system. Agreement between the reference assay and other measures was evaluated by the Bland–Altman method. Results Compared with the blood loss measured by the reference assay (470 ± 296 mL), the colorimetric system (572 ± 334 mL) was more accurate than either visual estimation (928 ± 261 mL) or gravimetric measurement (822 ± 489 mL). The correlation between the assay method and the colorimetric system was more predictive (standardized coefficient = 0.951, adjusted R2 = 0.902) than either visual estimation (standardized coefficient = 0.700, adjusted R2 = 00.479) or the gravimetric determination (standardized coefficient = 0.564, adjusted R2 = 0.304). Conclusion During cesarean delivery, measuring blood loss using colorimetric image analysis is superior to visual estimation and a gravimetric method. Implementation of colorimetric analysis may enhance the ability of management protocols to improve clinical outcomes.
No takes yet. Share an insight, caveat, or question.
Doctorvaladan et al. (2017) conducted an observational in Cesarean delivery (n=50). Colorimetric blood loss measurement (Triton system) vs. Reference extraction assay was evaluated on Total blood loss per procedure (mL) and mean bias compared to reference assay (MD 102, 95% CI 72-132, p=<0.001). During cesarean delivery, a colorimetric system measured blood loss with significantly greater accuracy (mean bias 102 mL) compared to visual estimation (mean bias 458 mL) and gravimetric measurement (mean bias 352 mL).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: