Key result
Fractal-coated iridium electrodes provided comparable monophasic action potential sensing features to standard Ag/AgCl electrodes, with small surface area tips yielding the best results.
Why the study?
Do fractal-coated iridium electrodes provide comparable monophasic action potential recordings to standard Ag/AgCl electrodes in patients undergoing electrophysiological studies?
Cross-Sectional (n=18)
Do fractal-coated iridium electrodes provide comparable monophasic action potential recordings to standard Ag/AgCl electrodes in patients undergoing electrophysiological studies?
Fractal-coated iridium electrodes, especially those with small surface area tips, offer comparable monophasic action potential sensing to standard Ag/AgCl electrodes.
MAP recording with fractal-coated iridium electrodes is feasible in EP studies; leaves open clinical utility for arrhythmia mapping.
The present study was designed to evaluate the feasibility of the recording of monophasic action potentials (MAP) with fractal-coated iridium electrodes in a clinical setting. In 18 patients who underwent an electrophysiological study for various arrhythmias, we performed MAP recordings with both 1.3-mm2 and 6-mm2 tip surface area fractal-coated iridium and standard silver--silver chloride (Ag/AgCl) electrodes in the high right atrium and two ventricular positions. Amplitude and MAP duration at 90%, 50%, and 25% of repolarization were calculated during steady-state pacing at 600, 500, and 400 ms cycle lengths with extrastimuli application. Morphology comparisons of MAP signals recorded with both types of electrodes were performed by regression analysis using 5% of the repolarization segments of the MAP trajectory. Differences between MAP duration at 90%, 50%, and 25% of repolarization recorded with fractal-coated and Ag/AgCl electrodes were statistically insignificant. Amplitude values recorded with 6-mm2 tip electrodes were significantly smaller than those recorded with Ag/AgCl electrodes for all comparisons. During steady-state pacing, the correlation coefficients between Ag/AgCl and fractal-coated 1.3-mm2 and 6-mm2 tip electrodes were within the range of 0.93-0.999 and 0.87-0.999, respectively. The correlation of MAP amplitude and duration at 90%, 50%, and 25% of repolarization following the extrastimulus S2, recorded with both types of electrodes, was significantly weaker for right atrial recordings (r value range 0.78-0.92) as compared to ventricular recordings (r value range 0.92-0.99). The MAP sensing features of fractal-coated iridium and Ag/AgCl electrodes are comparable. The best results for recording of MAPs with fractal-coated electrodes can be achieved with small surface area tip electrodes.
No takes yet. Share an insight, caveat, or question.
Zrenner et al. (2000) conducted a cross-sectional in various arrhythmias (n=18). Fractal-coated iridium electrodes vs. Standard silver-silver chloride (Ag/AgCl) electrodes was evaluated on Monophasic action potential duration and amplitude. Fractal-coated iridium electrodes provided comparable monophasic action potential sensing features to standard Ag/AgCl electrodes, with small surface area tips yielding the best results.