Key result
Impulse oscillometry measurements using the MasterScreen-IOS yielded significantly higher reactance values, resonant frequency, and area under the reactance curve compared with the MostGraph-02.
Why the study?
Limited data exist regarding both the absolute values and the percentage of predicted values of impulse oscillometry when measured using different devices.
Are impulse oscillometry measurements interchangeable between the MasterScreen-IOS and MostGraph-02 devices in patients with chronic respiratory diseases?
Cross-Sectional (n=54)
Are impulse oscillometry measurements interchangeable between the MasterScreen-IOS and MostGraph-02 devices in patients with chronic respiratory diseases?
Impulse oscillometry measurements from MasterScreen-IOS and MostGraph-02 devices show significant differences and should not be considered interchangeable for longitudinal assessments.
Device-specific IOS values may not be interchangeable in chronic respiratory disease; leaves open standardization across platforms.
BACKGROUND: Limited data exist regarding both the absolute values and the percentage of predicted values of impulse oscillometry (IOS) when measured using different devices. We aimed to compare IOS values and their percentages of predicted values obtained from the MasterScreen-IOS and the MostGraph-02 in individuals with chronic respiratory diseases. METHODS: test was used to compare IOS values derived from the two devices. Correlations between IOS values obtained from the two devices were analyzed using Pearson's correlation and Bland-Altman analysis was employed to determine the limits of agreement between the IOS values derived from these devices. RESULTS: A total of 54 subjects (36 with asthma and 18 with COPD) with a mean age of 65.7 ± 10.8 years, of whom 33 (61.1%) were male, were included. The reactance values, including the absolute values and percent predicted values for resonant frequency (Fres) and the area under the reactance curve between 5 Hz and the resonant frequency (AX), were significantly higher with the MasterScreen-IOS compared with the MostGraph-02. Discrepancies in IOS measurements, particularly in percent predicted values for heterogeneity of resistance between R5 and R20 (R5-R20), Fres, and AX, were observed when comparing the two devices. CONCLUSIONS: Significant differences in IOS reactance values were found between the MasterScreen-IOS and MostGraph-02. Device-specific reference equations may contribute to differences in percent predicted values. Therefore, IOS measurements from these devices should be interpreted cautiously and not considered interchangeable for longitudinal assessments.
No takes yet. Share an insight, caveat, or question.
Liwsrisakun et al. (2025) conducted a cross-sectional in Chronic respiratory diseases (asthma and COPD) (n=54). MasterScreen-IOS vs. MostGraph-02 was evaluated on Impulse oscillometry (IOS) values and their percentages of predicted values. Impulse oscillometry measurements using the MasterScreen-IOS yielded significantly higher reactance values, resonant frequency, and area under the reactance curve compared with the MostGraph-02.