Key result
A novel calibration-free method for computing arterial oxygen saturation based on a light attenuation model was developed and experimentally validated to be efficacious.
A novel analytical method for computing SpO2 eliminates the need for empirical calibration curves, making it independent of patient and sensor parameters.
Novel calibration-free SpO2 method feasible in humans; leaves open prospective validation before clinical adoption.
In present-day pulse oximeters, oxygen saturation in arterial blood( SpO₂)is computed by utilizing an empirical relationship extracted from a calibration curve. The calibration curve is obtained by curve-fitting data acquired from volunteers. A novel method of computation ofSpO₂that does not require the use of a calibration curve is presented in this paper. Based on a model for the attenuation of light through skin, tissue, bone, and blood, suitable processing steps are identified so that the analytical expression derived for the estimation ofSpO₂becomes free of not only patient but sensor-dependent parameters as well. The experimental results presented in this paper establish the efficacy of the proposed method.
No takes yet. Share an insight, caveat, or question.
Reddy et al. (2009) studied this question. Calibration-free method of computation of SpO2 vs. Standard calibration curve method was evaluated on Estimation of SpO2. A novel calibration-free method for computing arterial oxygen saturation based on a light attenuation model was developed and experimentally validated to be efficacious.