Key result
The HARPOON beating heart mitral valve repair system achieved successful cord implantation with mitral regurgitation reduction to ≤moderate in 91% of patients at 30 days.
Why the study?
The study was conducted to evaluate the safety and performance of a novel beating heart procedure for echocardiographic-guided implantation of ePTFE artificial cords in degenerative mitral regurgitation.
Does the HARPOON beating heart mitral valve repair system safely and effectively reduce mitral regurgitation in patients with degenerative mitral regurgitation?
Does the HARPOON beating heart mitral valve repair system safely and effectively reduce mitral regurgitation in patients with degenerative mitral regurgitation?
The HARPOON beating heart mitral valve repair system demonstrates high technical success and favorable 1-year clinical and echocardiographic outcomes in patients with degenerative mitral regurgitation.
May support transapical mitral repair feasibility in degenerative MR; leaves open need for RCTs on durability versus surgery.
OBJECTIVES: The objective of this study was to evaluate the safety and performance of a novel, beating heart procedure that enables echocardiographic-guided beating heart implantation of expanded polytetrafluoroethylene (ePTFE) artificial cords on the posterior mitral leaflet of patients with degenerative mitral regurgitation. METHODS: Two prospective multicentre studies enrolled 13 (first-in-human) and 52 subjects, respectively. Patients were treated with the HARPOON beating heart mitral valve repair system. The primary (30-day) end point was successful implantation of cord(s) with mitral regurgitation reduction to ≤moderate. An independent core laboratory analysed echocardiograms. RESULTS: Of 65 patients enrolled, 62 (95%) achieved technical success, 2 patients required conversion to open surgery and 1 procedure was terminated. The primary end point was met in 59/65 (91%) patients. Among the 62 treated patients, the mean procedural time was 2.1 ± 0.5 h. Through discharge, there were no deaths, strokes or renal failure events. At 1 year, 2 of the 62 patients died (3%) and 8 (13%) others required reoperations. At 1 year, 98% of the patients with HARPOON cords were in New York Heart Association class I or II, and mitral regurgitation was none/trace in 52% (n = 27), mild in 23% (n = 12), moderate in 23% (n = 12) and severe in 2% (n = 1). Favourable cardiac remodelling outcomes at 1 year included decreased end-diastolic left ventricular volume (153 ± 41 to 119 ± 28 ml) and diameter (53 ± 5 to 47 ± 6 mm), and the mean transmitral gradient was 1.4 ± 0.7 mmHg. CONCLUSIONS: This initial clinical experience with the HARPOON beating heart mitral valve repair system demonstrates encouraging early safety and performance. CLINICAL REGISTRATION NUMBERS: NCT02432196 and NCT02768870.
No takes yet. Share an insight, caveat, or question.
Gammie et al. (2020) studied Degenerative mitral regurgitation (n=65). HARPOON beating heart mitral valve repair system was evaluated on Successful implantation of cord(s) with mitral regurgitation reduction to ≤moderate at 30 days. The HARPOON beating heart mitral valve repair system achieved successful cord implantation with mitral regurgitation reduction to ≤moderate in 91% of patients at 30 days.