Key result
Point-of-care HbA1c testing linked to ~0.15-point lower patient HbA1c levels.
Cross-Sectional (n=22,778)
Yes
Answers are grounded in this paper and cite their sources.
Background It is not clear if point-of-care (POC) testing for hemoglobin A1c (HbA1c) is associated with glycemic control in type 2 diabetes. Methods In this cross-sectional study, we linked general practitioner (GP) data on 22,778 Norwegian type 2 diabetes patients to data from the Norwegian Organization for Quality Improvement of Laboratory Examinations. We used general and generalized linear mixed models to investigate if GP offices' availability (yes/no) and analytical quality of HbA1c POC testing (average yearly "trueness score", 0-4), as well as frequency of participation in HbA1c external quality assurance (EQA) surveys, were associated with patients' HbA1c levels during 2014-2017. Results Twenty-eight out of 393 GP offices (7%) did not perform HbA1c POC testing. After adjusting for confounders, their patients had on average 0.15% higher HbA1c levels (95% confidence interval (0.04-0.27) (1.7 mmol/mol [0.5-2.9]). GP offices participating in one or two yearly HbA1c EQA surveys, rather than the maximum of four, had patients with on average 0.17% higher HbA1c levels (0.06, 0.28) (1.8 mmol/mol [0.6, 3.1]). For each unit increase in the GP offices' HbA1c POC analytical trueness score, the patients' HbA1c levels were lower by 0.04% HbA1c (-0.09, -0.001) (-0.5 mmol/mol [-1.0, -0.01]). Conclusions Novel use of validated patient data in combination with laboratory EQA data showed that patients consulting GPs in offices that perform HbA1c POC testing, participate in HbA1c EQA surveys, and maintain good analytical quality have lower HbA1c levels. Accurate HbA1c POC results, available during consultations, may improve diabetes care.
No takes yet. Share an insight, caveat, or question.
Tollånes et al. (2020) conducted a cross-sectional in Type 2 diabetes (n=22,778). Absence of HbA1c point-of-care (POC) testing vs. Availability of HbA1c point-of-care testing was evaluated on Patients' HbA1c levels (MD 0.15%, 95% CI 0.04-0.27). Patients of general practitioners without HbA1c point-of-care testing had on average 0.15% higher HbA1c levels (95% CI 0.04-0.27) compared to those with available testing.
Synapse has enriched 10 related papers. Consider them for comparative context: