A sensitive and reliable method using liquid chromatography-tandem mass spectrometry (LC-MS/MS) with an electrospray-positive ionization method was developed for the determination of aflatoxin M(1) in milk. This method includes simple extraction of the sample with acetonitrile by ultrasonic, separation on an MGIII-C(18) column using 0.01% formic acid buffer/acetonitrile (60 : 40, v/v) as mobile phase, and MS/MS detection using multiple reaction-monitoring mode. Average recoveries of aflatoxin M(1) from spiked samples at concentrations of 0.02 and 1 ng ml(-1) ranged from 77% to 94%, with a 6% relative standard deviation. The limit of detection and limit of quantification were 0.006 and 0.02 ng ml(-1), respectively. The standard curve was linear between 0.02 and 20.0 ng ml(-1). The recommended method is simple, rapid, specific and reliable for the routine monitoring of aflatoxin M(1) in milk.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2010) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: