MICs of fluconazole and amphotericin B were determined independently for 100 coded yeast isolates by each of six laboratories to determine reproducibility of results by using a colorimetric oxidation-reduction-based broth microdilution test. In addition, each site tested five quality control isolates on at least four different occasions during the study. Results agreed within a three-dilution range (mode +/- 1 log2 dilution) for 96.2% of fluconazole tests and 92.7% of amphotericin B tests. Agreement among tests with the quality control isolates was 99.4% with fluconazole and 98.6% with amphotericin B. These results indicate that the colorimetric microdilution method is reproducible among laboratories.
No takes yet. Share an insight, caveat, or question.
Pfaller et al. (1994) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: