Key result
A smart textile system (Myant) reliably measured heart rate in horses with a mean bias of 0.31 bpm (95% limits of agreement -1.99 to 2.61) and strong correlation (r=0.9993, p<0.0001) versus telemetry.
Why the study?
ECG, heart rate, and heart rate variability are essential for assessing equine cardiovascular health, but limited validation work has been conducted for emerging smart textiles.
Does a smart textile system accurately measure HR and HRV in horses compared to a gold-standard telemetric device?
Observational (n=7)
Does a smart textile system accurately measure HR and HRV in horses compared to a gold-standard telemetric device?
Mean Difference: 0.31 (95% CI -1.99–2.61)
p-value: p=<0.0001
A smart textile system provides reliable measurements of heart rate and heart rate variability in horses at rest and during submaximal exercise compared to a gold-standard telemetric device.
May support smart textile ECG validation in horses; leaves open reliability for clinical use pending larger studies.
Electrocardiograms (ECGs), and associated heart rate (HR) and heart rate variability (HRV) measurements, are essential in assessing equine cardiovascular health and fitness. Smart textiles have gained popularity, but limited validation work has been conducted. Therefore, the objective of this study was to compare HR and HRV data obtained using a smart textile system (Myant) to the gold-standard telemetric device (Televet). Simultaneous ECGs were obtained using both systems in seven horses at rest and during a submaximal exercise test. Bland–Altman tests were used to assess agreement between the two devices. Strong to perfect correlations without significant differences between the two devices were observed for all metrics assessed. During exercise, mean biases of 0.31 bpm (95% limits of agreement: −1.99 to 2.61) for HR, 1.43 ms (−11.48 to 14.33) for standard deviation of R-R intervals (SDRR), and 0.04 (−2.30 to 2.38) for the HRV triangular index (TI) were observed. A very strong correlation was found between the two devices for HR (r = 0.9993, p < 0.0001) and for HRV parameters (SDRR r = 0.8765, p < 0.0001; TI r = 0.8712, p < 0.0001). This study demonstrates that a smart textile system is reliable for assessment of HR and HRV of horses at rest and during submaximal exercise.
No takes yet. Share an insight, caveat, or question.
McCrae et al. (2023) conducted an observational in Equine cardiovascular health (n=7). Smart textile system (Myant) vs. Gold-standard telemetric device (Televet) was evaluated on Heart rate (HR) and heart rate variability (HRV) measurements (Mean bias 0.31 bpm, 95% CI -1.99 to 2.61, p=<0.0001). A smart textile system (Myant) reliably measured heart rate in horses with a mean bias of 0.31 bpm (95% limits of agreement -1.99 to 2.61) and strong correlation (r=0.9993, p<0.0001) versus telemetry.