Answers are grounded in this paper and cite their sources.
An electrothermal vaporisation (ETV) device was developed for inductively coupled plasma mass spectrometry by modifying a commercial graphite furnace unit. A detailed description of the modifications are presented. The new ETV interface is designed to increase sample transport efficiency by converting analyte vapour into micro-particles in the furnace. The characterisation and optimisation of the device are also discussed. Ten consecutive firings of a 30-pg sample of Pb resulted in relative standard deviations (RSD) of 4% for peak-height and 1.6% for peak-area measurements. The absolute detection limit, determined by comparing the response obtained for a 3-µl sample of a 10 ng ml–1 Pb solution in 1% HNO3 with that obtained for blanks containing 1% HNO3 only, was 10 and 14 fg, for a dwell time of 10 µs, for peak area and peak height, respectively. The linear dynamic range for Pb extends over three orders of magnitude, from 300 fg to 300 pg.
No takes yet. Share an insight, caveat, or question.
Shen et al. (1990) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: