Key result
Among users with an initial irregular heart rhythm detection, the frequency of a recurrent detection was 57.2% (95% CI, 53.1%-60.9%) at 3 months, and longer ECG monitoring increased AF confirmation.
Why the study?
Wrist-worn wearables can detect irregular heart rhythms using photoplethysmography, but ECG confirmation is required to diagnose AF.
Does longer electrocardiographic monitoring improve the diagnostic yield for confirming atrial fibrillation in individuals with an initial irregular heart rhythm detection on a wearable device?
Cohort (n=1,057)
Does longer electrocardiographic monitoring improve the diagnostic yield for confirming atrial fibrillation in individuals with an initial irregular heart rhythm detection on a wearable device?
Longer electrocardiographic monitoring periods significantly increase the likelihood of confirming atrial fibrillation following an initial irregular heart rhythm detection by a wrist-worn wearable device.
Recurrent IHRDs after wearable detection support longer ECG monitoring to raise AF yield; leaves open optimal duration and practice adoption.
BACKGROUND: Wrist-worn wearables can detect irregular heart rhythms using photoplethysmography, but ECGs are required to confirm atrial fibrillation (AF). We sought to determine the frequency of a recurrent irregular heart rhythm detection (IHRD; ≥30 minutes of an irregular rhythm), estimate the potential diagnostic yield of different electrocardiographic monitoring strategies for confirming AF, and identify predictors of recurrent IHRDs. METHODS: The Fitbit Heart Study enrolled wrist-worn photoplethysmography device users without diagnosed AF. Of 455 699 participants, 1057 who wore and returned a 1-week ECG patch monitor after receiving an IHRD were analyzed. Baseline clinical data, device-derived metrics, IHRDs during follow-up, and electrocardiographic patch data were used for analysis. RESULTS: A total of 570 (53.9%) participants were aged 40 to 64 years, 422 (39.9%) were aged ≥65 years, and 510 (48.2%) were women. Median follow-up after ECG patch initiation was 80 days (interquartile range, 45–122 days). The frequency of another IHRD was 57.2% (95% CI, 53.1%–60.9%) at 3 months. After an initial IHRD, the estimated diagnostic yield for AF with a 10-second ECG was 7.6% (95% CI, 6.2%–9.0%), twice-daily 30-second ECGs over 1 week 19.0% (95% CI, 16.7%–21.2%), 24-hour monitor 17.4% (95% CI, 15.5%–19.3%), 1-week monitor 32.2% (95% CI, 29.4%–35.0%), 2-week monitor 46.8% (95% CI, 42.7%–50.8%), and 4-week monitor 60.8% (95% CI, 56.5%–65.1%). The risk of a recurrent IHRD was greater with older age ( P <0.001), male sex ( P =0.001), vascular disease ( P =0.03), longer initial runs of consecutive IHRDs at detection ( P =0.02), and less nightly sleep ( P =0.03). CONCLUSIONS: Irregular heart rhythms are common after initial detection using a wrist-worn wearable device. Longer electrocardiographic monitoring periods increase the likelihood of confirming AF. REGISTRATION: URL: https://www.clinicaltrials.gov ; Unique identifier: NCT04380415.
No takes yet. Share an insight, caveat, or question.
Lubitz et al. (2025) conducted a cohort in Irregular heart rhythm / Atrial fibrillation (n=1,057). Wrist-worn photoplethysmography device and ECG patch monitoring was evaluated on Frequency of a recurrent irregular heart rhythm detection (IHRD) at 3 months (95% CI 53.1-60.9). Among users with an initial irregular heart rhythm detection, the frequency of a recurrent detection was 57.2% (95% CI, 53.1%-60.9%) at 3 months, and longer ECG monitoring increased AF confirmation.