This is the authors' abstract. We don't add key points for this paper.
We demonstrate the operation of a quadrature phase-shifted extrinsic Fabry-Perot fiber-optic sensor for the detection of the amplitude and the relative polarity of dynamically varying strain. Two laterally displaced single-mode fibers inserted within a hollow silica tube form the 90 degrees phase-shifted sensing system. A multimode fiber, placed in the tube facing the two fibers, acts as a reflector, thereby creating an air gap that acts as a Fabry-Perot cavity. A theoretical description of the sensor is given, and its operation as a dynamically varying strain sensor is described. Strain sensitivities of 5.54 degrees phase shift/microstrain cm(-1) are obtained.
No takes yet. Share an insight, caveat, or question.
Murphy et al. (1991) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: