The determination of phosphorus and sulphur at trace levels is not possible by conventional atmospheric pressure plasma mass spectrometry because of high background signals at m/z= 31 (14N16OH+) and m/z= 32 (16O2+). The reduced-pressure plasmas decrease the background at the major isotopes of phosphorus and sulphur such that sub-nanogram detection is possible. The use of reduced-pressure helium and nitrogen microwave-induced plasmas for the determination of phosphorus and sulphur is discussed. With a helium plasma, the detection limit for phosphorus in triethyl phosphite ranges from 1 to 90 ng depending on the amount of torch cooling. Using a nitrogen plasma, sub-nanogram detection limits for phosphorus and sulphur in two organophosphorus pesticides were obtained.
No takes yet. Share an insight, caveat, or question.
Story et al. (1990) studied this question.