We describe a bent-crystal spectrometer adapted to measure x-ray emission resulting from core-level excitation of gas-phase molecules in the 0.8-8 keV energy range. The spectrometer is based on the Johann principle, and uses a microfocused photon beam to provide high-resolution (resolving power of approximately 7500). A gas cell was designed to hold a high-pressure (300 mbar) sample of gas while maintaining a high vacuum (10(-9) mbar) in the chamber. The cell was designed to optimize the counting rate (2000 cts/s at the maximum of the Cl Kalpha emission line), while minimizing self-absorption. Example of the Kalpha emission lines of CH(3)Cl molecules is presented to illustrate the capabilities of this new instrument.
No takes yet. Share an insight, caveat, or question.
Journel et al. (2009) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: