When the radiation source in a Michelson interferometer is placed in front of or behind the focal plane of the collimator, distortions arise in the spectral line shapes. The appearance and behavior of these distortions in a cube-corner interferometer are treated in this work. We shall also present a simple and fast method to calculate the true line position of a distorted line, and an easy rule of thumb to connect the off-focus shift and the amount of distortion. Calculation of the phase-error curves in the signal domain is treated as well. This calculation gives a possibility of correcting the distorted interferogram.
No takes yet. Share an insight, caveat, or question.
Saarinen et al. (1992) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: