Why the study?
Does a grid correction technique eliminate reconstruction artifacts in 3-D reconstructions of coronary bypass grafts from angiographic views?
Does a grid correction technique eliminate reconstruction artifacts in 3-D reconstructions of coronary bypass grafts from angiographic views?
Implementing a grid correction technique eliminates image distortion artifacts in 3-D reconstructions of coronary bypass grafts from X-ray angiograms.
May improve 3D graft reconstructions from angiograms; leaves open clinical validation in prospective human studies.
Three-dimensional (3-D) reconstructions of coronary bypass grafts performed from X-ray angiographic images may become increasingly important for the investigation of damaging mechanical stresses imposed to these vessels by the cyclic movement of the heart. Contrary to what we had experienced with coronary arteries, appreciable reconstruction artifacts frequently occur with grafts. In order to verify the hypothesis that those are caused by distortions present in the angiographic images (acquired with image intensifiers), we have implemented a grid correction technique in our 3-D reconstruction method and studied its efficiency with phantom experiments. In this article, the nature of the encountered artifacts and the way in which the dewarping correction eliminates them are illustrated by a phantom experiment and by the reconstruction of a real coronary bypass vein graft.