Measuring the intraocular pressure (IOP) in children can be difficult. We compared the accuracy of IOP measurements in pediatric patients using three tonometers: Perkins, Schiötz, and Tono-pen. Seventy-seven children (132 eyes; average age, 1 year, 7 months; range, 1 month to 60 months), mainly afflicted with retinopathy of prematurity (107 eyes), were evaluated. A good correlation (r = 0.867) was found between the Tono-pen and the Perkins tonometers. No statistical difference was found between the mean difference of IOP values obtained with the Tono-pen and the Perkins (P > .05). The Schiötz measurements were significantly higher than those obtained with the Perkins and the Tono-pen tonometers (P < .05). We concluded that the Tono-pen tonometer has a higher correlation coefficient, comparable to the Perkins tonometer, and can be used reliably to assess IOP in pediatric patients.
No takes yet. Share an insight, caveat, or question.
Bordon et al. (1995) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: