In magnetic resonance imaging (MRI) magnets, field shimming technique is, in general, necessary to increase the field homogeneity in imaging region to an acceptable level after manufacturing. This paper presents a novel passive shimming method using permanently magnetized shimming pieces. In the method, the shimming design is modeled as an integer programming problem, and the distribution of shimming pieces is optimized. Through an application to a 0.3-T permanent magnet MRI system, the optimally designed shimming pieces are proven to improve the field homogeneity from 1456 to 784 ppm after just one iteration of shimming courses is carried out.