Answers are grounded in this paper and cite their sources.
A design of STM which does not use vacuum internal vibration isolation and may be cooled to liquid helium temperatures is described. Tunneling current characteristics underlying STM operations are discussed. A model is presented to explain the anomalously low (<1 eV) tunneling barrier heights often observed. On the basis of this model we suggest criteria for obtaining reliable STM images.
No takes yet. Share an insight, caveat, or question.
Coombs et al. (1986) studied this question.