Why the study?
Do electroanatomic mapping systems and novel overlay guidance software provide accurate spatial and point localization compared to a ground truth phantom?
Do electroanatomic mapping systems and novel overlay guidance software provide accurate spatial and point localization compared to a ground truth phantom?
In a phantom model, CARTO3, EnSite Velocity, and a novel overlay guidance software all demonstrated high spatial accuracy with offsets below the clinically relevant threshold of 3 mm.
Answers are grounded in this paper and cite their sources.
Phantom accuracy is reassuring but does not support clinical adoption; leaves open in vivo validation of overlay guidance.
BACKGROUND: With increasing complexity in electrophysiology (EP) procedures, the use of electroanatomic mapping systems (EAMS) as a supplement to fluoroscopy has become common practice. This is the first study that evaluates spatial and point localization accuracy for 2 current EAMS, CARTO3(®) (Biosense Webster, Diamond Bar, CA, USA) and EnSite Velocity(®) (St. Jude Medical Inc., St. Paul, MN, USA), and for a novel overlay guidance (OG) software (Siemens AG, Forchheim, Germany) in a phantom experiment. METHODS AND RESULTS: A C-arm CT scan was performed on an acrylic phantom containing holes and location markers. Spatial accuracy was assessed for each system using distance measurements involving known markers inside the phantom and properly placed catheters. Anatomical maps of the phantom were acquired by each EAMS, whereas the 3D-based OG software superimposed an overlay image of the phantom, segmented from the C-arm CT data set, onto biplane fluoroscopy. Registration processes and landmark measurements quantitatively assessed the spatial accuracy of each technology with respect to the ground truth phantom. Point localization performance was 0.49 ± 0.25 mm in OG, 0.46 ± 0.17 mm in CARTO3(®) and 0.79 ± 0.83 mm in EnSite(®) . The registration offset between virtual visualization and reality was 1.10 ± 0.52 mm in OG, 1.62 ± 0.77 mm in CARTO3(®) and 2.02 ± 1.21 mm in EnSite(®) . The offset to phantom C-arm CT landmark measurements was 0.30 ± 0.26 mm in OG, 0.24 ± 0.21 mm in CARTO3(®) and 1.32 ± 0.98 mm in EnSite(®) . CONCLUSIONS: Each of the evaluated EP guidance systems showed a high level of accuracy; the observed offsets between the virtual 3D visualization and the real phantom were below a clinically relevant threshold of 3 mm.
No takes yet. Share an insight, caveat, or question.
Bourier et al. (2013) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: