A 96-well solid-phase extraction (SPE) system is used to rapidly prepare human urine samples for high-throughput quantitative analysis of two steroids, equilenin and progesterone, by liquid chromatography-tandem mass spectrometry using deuterated estrone as the internal standard. We define high-throughput here as analysis of 384 samples in a 24 h period. A total of 384 samples and standards were extracted by an individual in one day and subsequently analyzed within a 24 h period. The inter- and intratray accuracy and precision obtained over the course of these injections was within 8% coefficient of variation when analyzed by atmospheric pressure chemical ionization mass spectrometry using positive ion detection. A semiautomated sample processing workstation was used to add internal standard and then process 96 samples at a time. The recovery of the analytes from the SPE was approximately 85%. The accuracy and precision obtained was comparable to that ordinarily obtained using manual sample preparation techniques.