We have developed a wearable airbag that incorporates a fall-detection system that uses both acceleration and angular velocity signals to trigger inflation of the airbag. The fall-detection algorithm was devised using a thresholding technique with an accelerometer and gyro sensor. Sixteen subjects mimicked falls, and their acceleration waveforms were monitored. Then, we developed a fall-detection algorithm that could detect signals 300 ms before the fall. This signal was used as a trigger to inflate the airbag to a capacity of 2.4 L. Although the proposed system can help to prevent fall-related injuries, further development is needed to miniaturize the inflation system.
No takes yet. Share an insight, caveat, or question.
Tamura et al. (2009) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: