An HPLC method for the quantitation of paraquat in urine was applied to serum. Sample preparation consisted of ion-pair extraction on disposable cartridges of end-capped octadecyl silica. The paraquat thus extracted, was quantitated by HPLC using 1,1'-diethyl-4,4'-dipyridyl dichloride as an internal standard. Relatively small sample volumes (1 mL serum) were required. Within- and between-assay coefficients of variation for the HPLC technique were 2.9% and 4.0%, respectively, at low paraquat serum concentration (0.4 microgram/mL), and 2.4% and 3.2% at high paraquat serum concentration (4.0 micrograms/mL). The between-assay coefficient of variation for the total procedure, including the extraction, was 5.6% at low concentration (0.4 microgram/mL). Serum concentration levels down to 0.025 microgram/mL could be quantitated. The technique was checked for interference from muscle relaxants (pancuronium bromide and vecuronium bromide) and from anticoagulants (heparin and K2EDTA). No interference was observed.
No takes yet. Share an insight, caveat, or question.
Croes et al. (1993) studied this question.