A rapid and sensitive high-performance liquid chromatography and electrospray tandem mass spectrometry method was developed and validated for estimation of fulvestrant in rabbit plasma using liquid-liquid extraction. The separation and quantification of fulvestrant were achieved by reverse-phase chromatography on a Sunfire C18 column (50 x 2.1. i.d., 3.5 microm) with isocratic elution at a flow rate of 300 microL/min using norethistrone as an internal standard from 500 microL plasma sample. The method was validated over the concentration range from 0.092 to 16.937 ng/mL with a lower limit of detection of 0.023 ng/mL. The intra-day and inter-day accuracy and precision were within 10%. The recovery was 85 and 90% for fulvestrant and norethistrone respectively. The chromatographic run time was only 2.5 min.
No takes yet. Share an insight, caveat, or question.
Balaram et al. (2010) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: