We have developed an algorithm to automatically generate anatomy-based MLC fields. For each beam, a first field is adjusted to the projection of the target in a beam's eye view, allowing subsequent fields to be derived from this conformal field by removing the overlapping surface of each organ at risk, respectively. The projections are based on a surface sampling of the anatomical structures. On top of the MLC mechanical constraints, verification constraints are imposed on the MLC segments, in order to get reliable dosimetry using a commercial dose calculation engine. Thus, in each direction, the aperture's cross-section must be greater than a specified threshold, in our case 2 cm. Furthermore, junctions are not tolerated in order to avoid underdosage, for instance from the tongue-and-groove effect. The use of such MLC fields simplifies the verification process. The performance of the algorithm is illustrated for head and neck, thorax and prostate cases. Only a fraction of a second of CPU time is required to perform the segmentation for each beam.